2. If meter reads above ±4.4 then batteries are in good condition (C size)
2. Set MODE switch OPER and cheek the zero reading. Set the RANGE switch to least sensitive position (1,000 mS/m)
3. The tolerance for this test is ±1 mS/s. To adjust reading use the DC ZERO CONTROL, located under front panel. Remove battery pack to gain access to the controls.
4. Turn instrument off and attach receiver coil tube
2. Set MODE switch to OPER position and adjust the inphase meter reading to zero using the COARSE and FINE COMPENSATION controls. Tolerance +0.1 ppt.
3. To check the phase of the instrument, set the MODE switch to PHASE position. Note meter reading and rotate the COARSE control one step clockwise. If the conductivity meter reading remained the same (tolerance ±0.2), the phase is already correct. Return COARSE control to its original position.
4. If there is a difference in the readings, with the COARSE control in its original position, adjust the PHASE potentiometer about ¼ turn clockwise. Repeat the phase test. If the difference in readings has decreased, repeat procedure or if the difference has increased, the PHASE potentiometer should be rotated counter-clockwise.
5. N.B. Always remember to set the COARSE control back to its original position. This can be confirmed by checking that the inphase meter reads zero with the MODE switch set to OPER.
6. To check the sensitivity of the instrument, set MODE switch to COMP position and rotate COARSE control clockwise one step. The conductivity reading should change between 22 to 26 mS/m.
Vertical Dipole
Reading panel on top
Effective depth of exploration =6m
Horizontal Dipole
Reading panel is facing horizontally
Effective depth of exploration =3m
CONNECT THE LOGGER TO THE EM31 WHEN EM31 IS IN THE OPER POSITION and Press ENTER to start logging
REVIEW DATA MODE Press 6 (To review previous station information)
COMMENT Press 7 (enter comment then ENTER)
NEXT STN Press 5 (automatically skips a station)
NEW SEGMENT Press 1 (if you have to skip a large section)
NEW LINE Press 4 (Enter line no & Start stn. Enter '-'increment for reverse direction)
MEMORY Press 2 (See how much room is left on logger)
ESC Press ESC (Only use when finished logging)
Downloading the data from the Polycorder to your PC
1. Run DAT31 program on the PC and select COPY FILES FROM POLYCORDER (MOD-720)
2. Set Polycorder to MODE 5-2 and check that:
baud rate = 9600
data bits = 8
parity = N
mating call carriage return (CTRL 0 1 3)
3. Press ESC
4. On PC, Enter filename to store data: - Enter Hxxx file created in logger (CAPITALS), press RETURN
5. Set Polycorder to MODE 2-2 - Select corresponding file on polycorder press ENTER. And then press RETURN on PC.
6. Repeat procedure for Data files (D prefix)
7. Return to main menu (PC) and select CONVERT DL55 TO DAT31 FORMAT. Enter Hxxx filename and both the Hxxx and Dxxx files will be converted to xxx.G31 file format.
8. Use appropriate option to convert to Surfer format
Last updated by Tim Young (Tim.Young@GeoArch.demon.co.uk) 1st October 1998.